Fig 4.1 Scan Table Window

On Positions
Integration Time : Integration time for ON position. Typical value is 20 sec.
On Point Numbers : Numbers of start point and end point of ON positions. If you observe all ON positions defined in Scan Pattern, enter "1-total number of ON positions". You can observe some specified ON positions among the positions defined in Scan Pattern by using this parameter. Three sets can be defined.
absolute value : Absolute
value on the specified coordinate system.
Number of Sequences : Total number of scan per one ON position.
Interval : Number
of Sequence Pattern between intensity calibrations. Typical value
of the interval time of the calibration is 10-20 min. You should
decide this parameter considering the weather condition.
Auto Stop : You can stop your observation automatically, when the noise level becomes lower than a threshold that you define.
Fig 4.2 Frequency Switch Window

Throw Rate (Synthe 1 - 4) : Throw rate of frequency. The throw rate can be defined for 4 synthesizers individually.Modulation Rate :
time per phase : Integration time of one phase in one cycle.* Since the integration time unit is 2 (single) or 4 (double) times time per phase, Integration Time (common parameter) of ON position have to be a multiple of the unit.Switching mode :
single : Frequency of local oscillator is switched to one side (upper or lower).
double : Frequency of local oscillator is switched to both side (upper and lower).
switch rate : If you use only AOSs, select 0.25Hz (time per phase = 2.0s).Specifications of frequency switching is here
Fig 4.3 Beam Switch Window
Switching Mode :
Beam Separation :
Switching Frequency :

Ref. Array : Select a spectromer to measure rms noise level.
RMS Limit : Threshold value of rms noise level to stop observation.
Mode :POINT : If you observe only one ON position, use this mode.
MAP : For mapping observations, use this mode.No. of Ref. ON Poistion (MAP) : For MAP mode, you have to define an ON position as reference point to measure rms noise level.
Fig 4.5 Strip Scan Window
Coordinate System : Coordinate system for ON positions.Position Angle : Position angle of a strip of ON positions on the specified coordinate system. (+ : CCW)
Start Position : Coordinate of the start position. The origin is the coordinate of a source defined in Source Table.
Grid Spacing : Grid spacing of ON positions. Scan direction is defined by +/ -.
Grid Number : Number of ON position.
![]()
Fig 4.6 Scan pattern of the above Scan Table.
Fig 4.7 Map (single-grid) Window
Coordinate System : Coordinate system for ON positions.
Position Angle : Position angle of a map on the specified coordinate system. (+ : CCW)
Start Corner : Coordinate of the start position. The origin is the coordinate of a source defined in Source Table.
Grid Spacing : Grid spacing of X and Y directions. Scan direction is defined by +/-.
Grid Number : Grid numbers along X and Y axes. Total number of ON positions must be less than 10000.
Initial Direction : Initial direction of scan.
Fig 4.8 Scan pattern of the above Scan Table.
Fig 4.9 Map (multi-grid) Window
Coordinate System : Coordinate system for ON positions.
Position Angle : Position angle of a map on the specified coordinate system. (+ : CCW)
Large Grid : Large grid of mapping. After small grid points around one large grid point is scanned, the telescope moves to the next large grid point and starts scan of small grid points.
Start Corner : Coordinate of the start position. The origin is the coordinate of a source defined in Source Table.
Small Grid :
Grid Spacing : Grid spacing of X and Y direction. Scan direction is defined by +/-.
Grid Number : Grid numbers along X and Y axes. Total number of ON position must be less than 10000.
Initial Direction : Initial direction of scan.
Grid Spacing : Grid spacing along X and Y axes. Scan direction is defined by +/-.
Grid Number : Grid numbers along X and Y axes. Total number of ON position must be less than 10000.
Initial Direction : Initial direction of scan.
![]()
Fig 4.10 Scan pattern of the above Scan Table.
Fig 4.11 Random Position Window
Coordinate System : Coordinate system for ON positions.
Position Angle : Position angle of a map on the specified coordinate system. (+ : CCW)
Point (1) - (40) : Coordinate of ON positions. 40 ON positions can be defined.
Fig 4.12 Five Points Window
Observation Mode : For pointing observation, select pointing.
Coordinate System : Coordinate system of scan.
Position Angle : Position angle of scan on the specified coordinate system. (+ : CCW)
Number of Positions : Number of ON positions. Since the center is observed twice in one cycle, total number of ON positions is 6 for 5 points and 10 for 9 points.
Grid Spacing 1 : Position offset (of the inner points for 9 points) from the center.
Grid Spacing 2 (for 9 points) : Position offset of the outer points from the center.
- Fig 4.13 Scan pattern of the above Scan Table.
Fig 4.14 Spiral Scan Window
Coordinate System : Coordinate system for ON poisition.
Start Position : Coordinate of the start position. The origin is the coordinate of a source defined in Source Table.
Grid Spacing : Grid spacing of ON positions.
Specify Map Area by: You can define map size by length or number of grid.
Area Side Length : Length of one side of map. Grid number is N = (INT)(Area Side Length / Grid Spacing). If N is even, N = (INT)(Area Side Length / Grid Spacing) + 1.Initial Direction : Initial direction of scan.Side Grid Number : Grid number of one side of map. If you enter an even number, it is added by 1.
Position Angle : Position angle of initial direction of scan. (+ : CCW)Spiral Direction : Direction of spiral scan.
![]()
Fig 4.15 Scan pattern of the above Scan Table.
4.1.3 Tips
OFF1,
ON1,
ON2,
OFF1,
ON1,
ON2,
OFF1,
ON3,
ON4,
OFF1,
ON3,
ON4
OFF1,
ON1,
ON2,
OFF1,
ON3,
ON4,
OFF1,
ON1,
ON2,
OFF1,
ON3,
ON4
e.g.3) On Point Numbers : 2-6
Sequence Pattern
: 1**2**
Number of Sequence
: 2
OFF1,
ON2,
ON3,
OFF2,
ON4,
ON5,
OFF1,
ON6,
ON2,
OFF2,
ON3,
ON4,
OFF1,
ON5,
ON6